ENVAIR units have certificates. The microbiological safety workbenches are designed and constructed to fulfil the requirements of the European standard EN 12469:2000 and have TÜV certificates. ENVAIR microbiological safety workbenches are worldwide adopted for product, personnel and environment protection. The cabinets are especially suitable for handling dangerous materials which may be potentially harmful to humans and/or animals.

Features: ENVAIR eco safe Basic Mini

  • Energy consumption <281 W
  • Low noise level <56 dB(A)
  • Compact design for small scale applications
  • Hinged front visor for optimal cleaning
  • Stainless steel rear wall and work surface
  • Side walls of safety glass
  • Multi-function display with PLC control

Main applications

  • Microbiology
  • Virology
  • Haematology
  • Cell cultures
  • Genetics

Operation principle

  • 70 % recirculating air via main HEPA Filter H14
  • 30 % via exhaust HEPA Filter H14

Standard equipment

  • Hinged front visor for optimal cleaning
  • 1 electric socket
Model Type ENVAIR eco safe Basic Mini 0.7 m
Standards EN12469, EN61010-1
Overall Dimensions (W×H×D, mm) 830 × 980 × 650
Useful Dimensions (W×H×D, mm) 732 × 600 × 380
Inflow (m/s) 0.45
Downflow (m/s) 0.35
Electrical Data 1Ph+E – 230 V 50 hz

Support stands

Support stand with castor wheels

  • Two standard heights available: 750 mm or 900 mm
  • Durable polyurethane wheels with 360° rotation
  • Break system at the front wheels

Support stand with adjustable feet

  • Two standard heights available: 750 mm or 900 mm

Cabinet accessories

Electric socket and media connections

  • Electric sockets, Europe / Worldwide
  • Media connections (Compressed air, Gas, Vacuum, etc.)
  • Europe / Worldwide model DIN 12898, DIN 12919, DIN 3537

UV Lamp

  • Controlled by an automatic UV-timer via the microprocessor supported control panel.
  • Radiation of 253.7 Nm allows high-efficient decontamination
  • The lamp is mounted outside the view line of the user for good and safe radiation of the inner surfaces of the work zone.

Attention: Please observe the safety measures for your personnel when using the UV lamp. A UV lamp is not a substitution for good and effective cleaning and the effective lifetime is limited normally to 12 months, depending on the total number of operation hours. Replace the UV lamp every 12 months.

Ergonomic foot rest

  • Inclined, supports a good posture
  • Adjustable height
  • Anti-slide coating, chemical resistance

Ergonomic laboratory chair

  • Laboratory-suitable construction
  • Corresponds to cleanroom class 100
  • Alcohol-resistant PVC material
  • Adjustable height 395-490 mm

Microbiological Safety Workbench ENVAIR eco safe Basic Mini