Microbiological Safety Workbenches
ENVAIR eco safe Comfort Plus

ENVAIR units have certificates. The microbiological safety cabinets are designed and constructed to fulfil the requirements of the European standard EN 12469:2000 and have TÜV certificates. The ENVAIR microbiological safety workbenches are worldwide adopted for product, personnel and environment protection. The cabinets are especially suitable for handling dangerous materials which may be potentially harmful to humans and/or animals.

Main applications

  • Microbiology
  • Virology
  • Haematology
  • Cell cultures
  • Genetics

Operation principle

  • 70 % recirculating air via main HEPA Filter H14
  • 30 % via exhaust HEPA Filter H14

Most economic and most silent safety workbench worldwide

Test winner 2012 of ETH Zürich

Model Type ENVAIR eco safe Comfort Plus 0.9 m ENVAIR eco safe Comfort Plus 1.2 m ENVAIR eco safe Comfort Plus 1.5 m ENVAIR eco safe Comfort Plus 1.8 m
Standards EN12469, EN61010-1 EN12469, EN61010-1 EN12469, EN61010-1 EN12469, EN61010-1
Overall Dimensions (W×H×D, mm) 1045 × 1500 × 855 1350 × 1500 × 855 1655 × 1500 × 855 1960 × 1500 x 855
Useful Dimensions (W×H×D, mm) 887 × 740 × 580 1192 × 740 × 580 1497 × 740 × 580 1802 × 740 × 580
Inflow (m/s) 0.45 0.45 0.45 0.45
Downflow (m/s) 0.25 0.25 0.25 0.25
Electrical Data 1Ph+E – 230 V 50 hz 1Ph+E – 230 V 50 hz 1Ph+E – 230 V 50 hz 1Ph+E – 230 V 50 hz

Support stands

Support stand with castor wheels

  • Two standard heights available: 750mm or 900mm
  • Durable polyurethane wheels with 360° rotation
  • Break system at the front wheels

Support stand with adjustable feet

  • Two standard heights available: 750mm or 900mm

Cabinet accessories

Electric socket and media connections

  • Electric sockets, Europe / Worldwide
  • Media connections (Compressed air, Gas, Vacuum, etc.)
  • Europe / Worldwide model DIN 12898, DIN 12919, DIN 3537

IV suspension bar with hooks
An attached IV suspension bar offers on pharmaceutical applications an easy possibility to hang or store IV bags during the work course. Besides it increases the effectiveness in the work area.

  • Stainless steel IV bar with 6 hooks

UV Lamp

  • Controlled by an automatic UV-timer via the microprocessor supported control panel.
  • Radiation of 253.7 Nm allows high-efficient decontamination
  • The lamp is mounted outside the view line of the user for good and safe radiation of the inner surfaces of the work zone.

Attention: Please observe the safety measures for your personnel when using the UV lamp. A UV lamp is not a substitution for good and effective cleaning and the effective lifetime is limited normally to 12 months, depending on the total number of operation hours. Replace the UV lamp every 12 months.

Ergonomic foot rest

  • Inclined, supports a good posture
  • Adjustable height
  • Anti-slide coating, chemical resistance

Ergonomic laboratory chair

  • Laboratory-suitable construction
  • Corresponds to cleanroom class 100
  • Alcohol-resistant PVC material
  • Adjustable height 395-490 mm

Microbiological Safety Workbench ENVAIR eco safe Comfort Plus

Multi-function display

Pre-filter for Safety Workbench